Vertical Deflection Accuracy - Tektronix 442 Service Manual

Hide thumbs Also See for 442:
Table of Contents

Advertisement

AMPLITUDE
CALIBRATOR
Calibration Procedure—442 Service
Performance
Check
COUPLER
PROCEDURE
NOTE
The
following
procedure
is arranged
to be per-
formed
in its entirety,
starting
with
Step
1 and
proceeding in sequence to the end. This ensures the
correct 442 and test equipment settings and connéc-
tions for each step.
1.
Channel
1 and Channel 2 Deflection Accuracy
Check
a.
Connect the test equipment as shown in Figure 4-1
and
apply
an
unterminated
10 mV,
1 kHz
Amplitude
Calibrator signal to the CH 1 and CH 2 input connectors.
Set the CH 1 AC-GND-DC switch to DC and the CH 2 AC-
GND-DC switch to GND. In the remainder of this step, set
the CH 1 and CH 2 POSITION control (as appropriate) to
center the display within the graticule area.
b.
CHECK—Channel
1 deflection
accuracy
for the
settings listed in Table 4-2 within 3% (+20°C to +30°C).
Adjust
INTENSITY
and
all
POSITION
controls
as re-
quired. For best results in Steps 1 and 2, leave the LEVEL
control fully clockwise for a free-running display.
c.
Set:
Amplitude Calibrator
Output
10 mV
442
AC-GND-DC
(CH 1)
GND
AC-GND-DC
(CH 2)
DC
VERT
MODE
CH
2
POSITION
(CH 2)
For centered display
REV A MAY 1980
d.
CHECK—Channel
2 deflection
accuracy
for the
settings listed in Table 4-2 within 3% (+20°C to +30°C).
Set CH 2 POSITION
as required.
Table 4-2
VERTICAL
DEFLECTION
ACCURACY
VOLTS/DIV | Amplitude
Vertical
+%
Setting
Calibrator
Deflection
Tolerance
(1X Window)
Output
(Divisions) |
(Divisions)
——————
2m
10 mV
4.85 to 5.15
3.88 to 4.12
4.85 to 5.15
4.85 to 5.15
4.85 to 5.15
4.85 to 5.15
5m
20 mV
10m
50 mV
20m
TV
2
1V
2
10V
45