Appendix
C:
Performance Verification
Table A-8: Deflection Accuracy Limits
VOLTSIDIV
Switch
Setting
Standard
Amplitude
Signal
Accuracy
Limits
(Divisions)
c. Move the cable from the CH 1 OR X input connector to the CH 2 O R
Y input connector. Set the Vertical MODE switch to CH 2.
d. Repeat part b using the CH 2 controls.
Step 2: Check Store Deflection Accuracy
a. Set:
CH 2 VOLTSIDIV
STOREINON-STORE
Acquisition MODE
2
mV
STORE (button in)
AVERAGE
b. Set the generator to produce a five division standard amplitude
signal.
c. Use the CURSORS control and SELECT C1lC2 switch (push in the
CURSORS controls knob) to set one cursor at the bottom of the
square wave and the other cursor at the top of the square wave.
d. CHECK
-
Deflection accuracy is within the limits given in Table A-9
for each CH 2 VOLTSIDIV switch setting and corresponding stan-
dard-amplitude signal.
e. Move the cable from the CH 2 O R Y input connector to the CH 1 O R
X input connector. Set the Vertical MODE switch to CH 1.
f.
Repeat parts b and c using the CH 1 controls.