Performance Tests
C-14
Equipment Required (See Table C-1)
Calibration Generator (Item 1)
Leveled Sine-Wave Generator (Item 2)
Time-Mark Generator (Item 3)
50
Q
Coaxial Cable (Item 7)
50
Q
BNC Termination (Item 11)
Initial Control Settings
Vertical (CH 1 and CH 2)
POSITION ..................................... Midrange
MODE ............................................ CH 1
VOLTS/DIV ...................................
0.5
V
V ARiable ....................................... Off
AC-GND-DC .................................. DC
Horizontal
POSITION ..................................... Midrange
MODE ALT MAGN ....................... Off
SEC/DIV .........................................
0.05
µs
V ARiable ....................................... Off
Xl, XlO, X50
Magnifier ................
Xl
Trigger
HOLDOFF ...................................... MIN
LEVEL ............................................ Midrange
SLOPE ............................................ Positive Going
MODE ............................................ P-P AUTO
SOURCE ........................................ VERTical MODE
COUPLING .................................... AC