Table C-3 Settings For Timing Accuracy Checks - Tektronix 2212 User Manual

Digital storage & analog oscilloscope
Hide thumbs Also See for 2212:
Table of Contents

Advertisement

Performance Tests
Table C-3
Settings for Timing Accuracy Checks
SEC/DIV
Time-Mark Generator Setting
Switch Setting
Xl
(Normal)
XlO
Magnify
X50
Magnify
0.05 µs
50 ns
5
ns
=1
ns
0.1 µs
0.1 µs
10 ns
=2 ns
0.2 µs
0.2 µs
20 ns
=4 ns
0.5 µs
0.5 µs
50 ns
10
ns
1
µs
1
µs
0.1 µs
20 ns
2
µs
2
µs
0.2 µs
0.04µs
5
µs
5
µs
0.5 µs
0.1 µs
10
µs
10
µs
1
µs
0.2 µs
20
µs
20 µs
2
µs
0.4 µs
50
µs
50 µs
5
µs
1
µs
0.1 ms
0.1 ms
10
µs
2
µs
0.2 ms
0.2 ms
20 µs
4
µs
0.5 ms
0.5 ms
50 µs
10
µs
1
ms
1
ms
0.1 ms
20 µs
2
ms
2
ms
0.2 ms
0.04ms
5
ms
5
ms
0.5 ms
0.1 ms
10
ms
10
ms
1
ms
0.2 ms
20 ms
20 ms
2
ms
0.4 ms
50 ms
50 ms
5
ms
1
ms
0.1
s
0.1
s
10
ms
2
ms
0.2
s
0.2
s
20 ms
4
ms
0.5
s
0.5
s
50 ms
10 ms
2212 User Manual