SpectraMax Paradigm Multi-Mode Detection Platform User Guide
8. Follow the on-screen instructions in the Instrument Unlocking Procedure wizard to
unlock the drawers and initialize the instrument.
After completing the Instrument Unlocking Procedure wizard, the instrument does an
initialization procedure that moves the optics and microplate drawers to home positions.
The green LED on the status panel turns on, the amber LED flashes and turns off, and then
the green LED remains on. You can install the desired detection cartridges in the instrument.
Connecting and Disconnecting a Gas Supply
A gas supply, such as a CO
SpectraMax Paradigm Multi-Mode Detection Platform. This is useful when reading plates as
part of a cell-based assay in which a CO
cultures alive. The gas supply is not regulated or monitored by the instrument or software.
These procedures require the following tools and accessories:
Table 3-5: Required Tools and Accessories
Part Number
Not applicable
Not applicable
S MS 135 100
76
Note: If the instrument does not show in the Available Instruments list of the
Instrument Connection dialog, then click Refresh above the list.
, nitrogen, or other gas supply, can be connected to the
2
Description
Polyurethane tubing, outside diameter = 4.0 mm
Slot-head screwdriver
Gas inlet unlock tool (not provided with the instrument)
environment needs to be provided to keep cell
2
5014038 E
Need help?
Do you have a question about the SpectraMax Paradigm and is the answer not in the manual?
Questions and answers