Table A-3: Environmental Controls of the SpectraMax Paradigm Multi-Mode Detection
Platform
Item
Temperature range
Displayed
Accuracy
Uniformity
Gas inlet specification Maximum air supply pressure: 0.4 MPa
Gas tubing
specification
WARNING!
instrument is not air-tight, and so gas can escape into the atmosphere surrounding
the instrument. When using potentially toxic gas, always observe the applicable
cautionary procedures as defined by your safety officer to maintain a safe working
environment.
5014038 E
Description
LED-based detection cartridges:
4°C (7.2°F) above ambient to 45°C (113°F)
Xenon flash lamp based detection cartridges:
5°C (7.2°F) above ambient to 45°C (113°F)
1°C (1.8°F) increments
±1°C (1.8°F) at 37°C set point
±1.5°C (2.7°F)
Polyurethane tubing, outside diameter = 4.0 mm
 Use a compressed gas supply in a well-ventilated area. The
Appendix A: Instrument Specifications
169