3.3. Print Limits

This Chg Lab Limits sub-menu allows the operator to print out the current Laboratory Limits and
Morphology Flag values.
Changing Laboratory limits
Main menu
5- Setup
2- Chg Lab Limits
3- Prt Limits
Press the «Enter» Key
The current Laboratory Limits will automatically print out as shown in
printout, page 8"
Fig.5-1: Laboratory limits printout
Verify your entries. Check each parameter to make sure that you have entered the correct number
and placement of the decimal point.
PCT and PDW are not available in the United States. These parameters are strictly used
for research and investigational purposes Only.
8 - RAB042FEN - User Manual -
60
L I M I T S
DATE : 01/20/2002
TIME : 10:54
LOW
HIGH
WBC
: 3.5
10.0
RBC
: 3.80
5.80
HGB
: 11.0
16.5
HCT
: 35.0
50.0
MCV
: 80
97
MCH
: 26.5
33.5
MCHC
: 31.5
35.0
PLT
: 150
390
MPV
: 6.5
11.0
RDW
: 10.0
15.0
%LYM
: 17.0
48.0
%MON
: 4.0
10.0
%GRA
: 43.0
76.0
#LYM
: 1.2
3.2
#MON
: 0.3
0.8
#GRA
: 1.2
6.8
PLT Flags :
SCL%
: 8.0
SCH%
: 8.0
MIC%
: 8.0
WBC Flags :
L1%
: 8.0
M2%
: 8.0
G1%
: 15.0
G3%
: 8.0
ABX Micros
60
"Fig.5-1: Laboratory limits
3
3
10
/ mm
6
3
10
/ mm
g/dl
%
3
µm
pg
g/dl
3
3
10
/ mm
3
µm
%
%
%
%
3
3
10
/ mm
3
3
10
/ mm
3
3
10
/ mm