ABX Micros

6.1. Host Communications

4 different Output formats are available:
1-
Format ARGOS, No character identifiers
2-
Format R&D, No character identifiers
3-
Format ABX, with character identifiers
4-
TR OFF, Transmission Off.
Setting output format
Main Menu
5- Setup
6- Host options
1- Host Comm.
Select between the 4 above formats
Press the «Enter» key to accept that specific output format.

6.2. Baud Rate

This sub-menu allows the operator to select the rate of transmitted data to the host computer. There
are 5 different rates to select from:
1-
300
2-
1200
3-
2400
4-
4800
5-
9600
Setting Baud rate
Main Menu
5- Setup
6- Host options
2- Baud rate.
Select between the 5 above rate
Press the «Enter» key to accept that specific rate.

6.3. Transmission

This sub-menu will allow the operator to send the last sample results in memory to the Host com-
puter as a transmission acceptance verification. This function is mainly used to verify transmission
from the
ABX Micros
Setting Baud rate
Main Menu
5- Setup
6- Host options
3- Transmission
Press the «Enter» key to send the last sample results to the Host computer.
16 - RAB042FEN - User Manual -
60
60
to a host
computer.
ABX Micros
60