1.4. Instrument Cycles
This Service function provides the number of cycles the
menu will contain 3 cycles to view as indicated:
1 - Startup: the number of Startup cycles that the instrument has ran.
2 - Standby: the number of Standby cycles that the instrument has ran.
3 - CBC: the number of analysis cycles that the instrument has ran.
Viewing cycles number
Main menu
4- Service
6- Cycles
Use the «Up» and «Down» arrows
1.5. Technician Functions
This Service function allows the operator to work with HORIBA ABX Technical Support in resolving,
checking, and/or adjusting some technical operations on
This Service Menu is mainly used by the Engineers to make adjustment in the technical areas of the
analyzer.
This Menu can be accessed only by a special password. When troubleshooting your in-
strument in this area, call your local HORIBA ABX Technical Support Representative for
the special password and instructions in this menu area.
1.6. Automatic Cleaning
Provides an automatic Cleaning cycle to run at any time when the operator wishes to clean the
instrument.
Viewing cycle number
Main menu
4- Service
8- Autoclean
Press the «Enter» key
This Auto Clean cycle is a short cycle and may not provide enough cleaning action for
Fibrin clots in the apertures and/or chambers. If this is the case, a Concentrated Clean-
ing is highly recommended (see
1.7. Manual Probe Cleaning
Concerns:
◆Unblocking Sample probe
◆Replacing Sample probe
Instrument Response/Problem type
◆Low results on all parameters in any analysis mode
Maintenance
ABX Micros
ABX Micros
"1.3.6. Concentrated Cleaning, page
ABX Micros
& Troubleshooting
60
has run. This cycle
60.
11").
60
- User Manual - RAB042FEN - 13
Maintenance