Problems On All Parameters - HORIBA ABX Micros 60 User Manual

Hematology analyzer
Table of Contents

Advertisement

ABX Micros

2.3.1. Problems on all Parameters

Tab.6-2:
Problem Noted
Startup Failed
No results in analysis
Suspicious patient results
Alarm flags on most parameters
QC outside acceptable limits
Poor repeatability
Poor Chamber drainage
20 - RAB042FEN - User Manual -
60
Possible Cause
Out of Reagent(s)
Contaminated Reagent(s)
Possible dirty Apertures
Blocked sample probe
Possible defective Pre-amp on Main
Board
Possible dirty Apertures
Possible partial blocked sample probe Remove and flush out sample probe
Possible leakage from syringe blocks
Possible Fibrin clot in «T» fitting be-
low WBC chamber
Possible dirty Apertures
Possible Fibrin clot in «T» fitting be-
low WBC chamber
Possible dirty apertures
Possible leakage from syringe block
Possible Fibrin clot in «T» fitting be-
low WBC chamber
Poor calibration
Possible dirty apertures
Possible leakage from syringe block
Possible Fibrin clot in «T» fitting be-
low WBC chamber
Poor calibration
Possible Fibrin clot in «T» fitting be-
low WBC chamber
Possible Vacuum/Waste syringe
malfunction
ABX Micros
60
Corrective Action
Check and/or replace reagent(s) if
necessary
Check and/or replace reagent(s) if
necessary
Perform a concentrated cleaning
"1.3.6. Concentrated Cleaning,
page
11", re-run startup
Remove and flush out sample probe as
1.7. Manual
descibed in Section 6,
Probe Cleaning
, page 6-14
Call HORIBA ABX technical support for
Assistance
Perform a Concentrated cleaning
"1.3.6. Concentrated Cleaning,
page
11", re-run sample
Check syringe block for leakage and call
HORIBA ABX Technical support
Call HORIBA ABX Technical support
Perform a Concentrated cleaning
"1.3.6. Concentrated Cleaning,
page
11", re-run Sample
call HORIBA ABX Technical support
Perform a concentrated cleaning
"1.3.6. Concentrated Cleaning,
page
11", re-run sample
Check syringe block for leakage and call
HORIBA ABX Technical support
Call HORIBA ABX Technical support
Re-calibrate and rerun QC (See Section
2. Calibration
4,
, page 4-11)
Perform a concentrated cleaning
"1.3.6. Concentrated Cleaning,
page
11", re-run sample
Check syringe block for leakage and call
HORIBA ABX Technical support
Call HORIBA ABX Technical support
Call HORIBA ABX Technical support
Call HORIBA ABX Technical support
check syringe correct operation as de-
scribed in
"1.3.7. Mechanical Checks,
page 12"