Results Rejected; Results Comparable; Hgb Blank Reference - HORIBA ABX Micros 60 User Manual

Hematology analyzer
Table of Contents

Advertisement

:
Tab.3-2: Printed and host transmitted results
Parameters
3
3
WBC(x10
/mm
)
6
3
RBC(x10
/mm
)
3
3
PLT(x10
/mm
)
HGB>2g/dL
3
3
PLT(x10
/mm
)
HGB<2g/dL
HGB(g/dL)
HCT(%)
Whole Blood parameter results within visible range will still give a result value with a
«D» indication. These results require a dilution (or PRP analysis for PLT) of the sample.

6.2.2. Results Rejected

A «Reject» flag, shown by an asterisk (*) following the Parameters WBC, RBC, HCT, or PLT indicates
that the
ABX Micros
counts were outside the system precision limits for that specific parameter.
The results should be verified by re-running the sample.

6.2.3. Results Comparable

A reportable result shown by a Dollar sign ($) following the parameters WBC, RBC, HCT, or PLT indi-
cates that the
ABX Micros
out of three counts were within the system precision limits for that specific parameter. The results
can be reported, but the parameter should be monitored for the next sample analyzed.

6.2.4. HGB Blank Reference

A suspicious flag, shown by an Exclamation (!) located next to the Hemoglobin result shows that
the HGB Blank carried out during the analysis differs from the previous cycle's HGB Blank. This (!)
means that both HGB blanks were outside the Instrument precision limits.
This result can be reported, but the parameter should be monitored for the next sample analyzed.
The MCH and MCHC may also be affected by this (!) flag depending on the severity of the results.
If the Exclamation (!) flag occurs more than 3 consecutive sample runs, troubleshoot
the HGB parameter in the following section of this manual. Section 6,
2.
Troubleshooting, page 6-18.
Linearity
Visible Range
limits
«result»
«result»+ «D»
«result»
«result»+ «D»
«result»
«result»+ «D»
«result»
«result»+ «D»
«result»
«result»+ «D»
«result»
«result»+ «D»
60
has analyzed that parameter for a maximum of 3 counts. All 3
60
has analyzed that parameter for a maximum of 3 counts. Two
ABX Micros
Workflow
> Visible Range
--.-- + D
--.-- + O
--.-- + D
--.-- + O
--.-- + D
--.-- + O
--.-- + D
--.-- + O
--.-- + D
--.-- + O
--.-- + D
--.-- + O
60
- User Manual - RAB042FEN - 15
Flags

Advertisement

Table of Contents
loading

Table of Contents