Now take all 3 levels of controls, LOW, NORMAL, HIGH, and run them 1 time each as a regular
patient analysis. When the cycles are complete, note only the RDW results and compare them to the
control Assay values for RDW, on the Assay sheet. Verify that the RDW results are somewhat close to
the Mean Assay values and within the ranges as specified for all 3 Levels.
8-
The RDW calibration is now complete. Be sure to monitor your RDW results and verify that they
fall within your patient population.
9-
Verify that Normal RDW results will be within the established ranges set in the setup menu. See
Section 5,
Certain Pathological conditions will affect the RDW results. See (See Section 1,
4.3. Known Interfering

2.2.12. Calibration coefficients

The operator has the option of either «Editing» the current calibration coefficients or «Printing» the
current calibration coefficients.
The ABX Micros
tained. Should any variation of Quality Control results outside the Assayed ranges occur
after calibration, it is Highly suggested that you contact your local HORIBA ABX Tech-
nical Support Representative before Manually editing the calibration coefficients!

2.2.13. Changing Calibration Coefficients

Editing calibration coefficients
Main menu
3- Calibration
2- Coefficients
1- Calib. Coeff.
Enter password 123 or the one defined by the operator in the setup menu. See Section 5,
4.2. Changing (the User) Password
Press «Enter» key
Press «Up» or «Down» arrow keys to select the coefficient to be edited.
Press «Enter» key and enter new coefficient derived from using the below formula:
Press «Enter» key to accept the coefficient.
Continue to use the «Up» or «Down» arrow keys to select the next coefficient to be edited.
Use the same formula below to calculate the remaining coefficients.
Press the «Esc» key until you return to the Main Menu
◆New coefficient calculation
(Parameter target value/parameter mean value) x Current parameter coeff. = New parameter coeff.
After manually editing the calibration coefficients, it is highly recommended to run
Quality Control. Verify that all levels of control material are within their specified pa-
rameter ranges. Verify that there are no error flags (H, L, *, $, ! for HGB) associated
with all levels of Quality control results.
Calibration Coefficients are as listed:
18 - RAB042FEN - User Manual -
60
3. Laboratory Limits
, page 5-7.
Substances, page 1-14)
60
is an accurate and reliable instrument when properly main-
WBC
RBC
HGB
HCT
ABX Micros
, page 5-10