Replacing The Sample Probe - HORIBA ABX Micros 60 User Manual

Hematology analyzer
Table of Contents

Advertisement

11-
Hold the Sample probe mounting bracket on the upper right edge and move the sample probe
upwards until the tip of the sample probe is about 1 inch below the rinsing block.
12-
Press the «Startup» key and verify that the background results are within their limits.
13-
Re-run the Quality Controls to verify correct results.

1.7.2. Replacing the Sample Probe

Procedure:
1-
Enter the service Menu
Entering service menu
Main menu
4- Service
2-
Open the main door of the
3-
Now press the «Esc» key to open the tube holder door.
4-
Hold the sample probe mounting bracket on the upper right edge and move the sample probe
Downwards until you can see the top of the probe and sample tubing.
5-
Remove the tubing from the top of the sample probe.
6-
Remove the sample probe Retaining Clip. (Pull the clip out horizontally towards the front of the
analyzer).
7-
Hold the sample probe mounting bracket on the upper right edge and move the sample Upwards
until the top of the probe is just below the front panel.
8-
Very carefully pull the sample probe slightly out of its retaining slot.
9-
Now hold the sample probe around the retaining collar very carefully lift up the probe so that it
is free from the white needle guide.
The sample probe is fragile and may bend if not inserted into the needle guide correctly.
Be carefull when placing the probe. Try to maintain a vertical position of the probe when
inserting it into the white needle guide.
10-
Place the new sample probe in the white needle guide.
11-
Place the retaining collar of the probe into its slot.
12-
Hold the Sample probe mounting bracket on the upper right edge and move the sample probe
upwards until the tip of the sample probe is about 1 inch below the rinsing block.
13-
Re-connect the sample tubing to the top of the probe so that it is air tight.
14-
Hold the Sample probe mounting bracket on the upper right edge and move the sample probe
upwards until the tip of the sample probe is about 1 inch below the rinsing block.
15-
Press the «Startup» key and verify that the background results are within their limits.
16-
Re-run the Quality Controls to verify correct results.
Maintenance
ABX Micros
60
(see
"1.3.2. Cover dismantling, page
ABX Micros
& Troubleshooting
8").
60
- User Manual - RAB042FEN - 15