2. Daily Quality Control & Calibration verification
Before analyzing any patient blood samples, it is recommended that the Operator performs Quality
Control analysis on 3 levels of Control Blood Material, (Low, Normal, and High), to verify that the
ABX Micros
Please refer to Section 4,
General recommendations on calibration
The calibration on HORIBA ABX instruments is an exceptional procedure, which must be
carried out particularly in case of certain technical interventions (installation, mainte-
nance, service intervention). The calibration should not be carried out to compensate a
drift on a result due for example to a clogging of the instrument.
Running control blood
Mix the sample gently and thoroughly.
Remove the cap from the sample tube.
Place the sample beneath the sample probe.
Raise up the tube so that the sample probe lowers into the blood, and press the manual sam-
ple bar or press the «Startup» key.
The analysis cycle will begin.
If any results are out of their acceptable ranges, perform the following steps:
1-
Rerun the Control blood.
2-
Clean the system, see Section 6,
3-
Open a new vial of control material.
4-
Call your local HORIBA ABX Technical support representative before re-Calibrating the system.
4 - RAB042FEN - User Manual -
60
60
is performing within the specified ranges of the Quality control material.
1. Quality Control
, page 4-2 in order to perform the Control sample run.
1.3.6. Concentrated
ABX Micros
60
Cleaning, page 6-11. Re-run the sample.