Tab.6-6:
Error message
No printer error
Printer not selected error
Bad Date, Try again! error
Bad Time, Try again! error
Startup cycle not initiated
Bad Value... MINI : XXX, MAXI : XXX error
Minimum and Maximum values are displayed with the corresponding error message! Cor-
rect your value to be within the specified ranges given. If unable to be within the range
specified, call your local HORIBA ABX Technical Support Representative.

3. Menu/Overview

Tab.6-7: Menu «RESULTS»
1- RESULTS
Parameters 1
Pat. ID / Run #
Flags
Parameters 2
Tab.6-8: Menu «QC»
2- QC
1- Automatic
2- Analysis
3- Print targets
Maintenance
Corrective Action
Printer is Off Line, reset printer (See Section 5,
2.1. Reprint Results
Power On and/or reconnect printer to unit (See Section
2.1. Reprint Results
5,
Incorrect printer setup (See Section 5,
Selection
, page 5-5)
Use the Period ( . ) key between each group of numbers
when entering the date.
Possible wrong date format, check for the correct date
format, re-enter the date. Use the Period( . ) key be-
tween each group of numbers when entering the date.
Use the Period( . ) key between each group of numbers
when entering the time.
Press the «Startup» key to run a Startup.
◆ Enter correct value between (0 to 99,999) for sample
Run #
◆ Out of range Target value in Autocalibration. Enter
the correct target value
◆ Out of range run number in Autocalibration. Enter
the correct run number between (3 to 11)
◆ Cal coefficient out of range. Enter the coefficient
within the range specified in Calibration & Quality
Control section
QC Smart card menu
Select OP.
lot #
Select level (1- Low, 2- Normal, 3- High)
QC Smart card menu
Select OP.
lot #
Select level (1- Low, 2- Normal, 3- High)
ABX Micros
& Troubleshooting

Menu/Overview

, page 5-3)
, page 5-3)