1-
Insert the Q.C. Smart Card into the reader with a firm push until it «clicks» into place.
Printing targets
Main Menu
2- QC
3- Print targets
«Lot#» and «Exp Date» are displayed
Press «Enter» key
The printout will show Low limit and High limit for each parameter, for each assayed level of control.
HIGH, NORMAL,and LOW.
1.5. Q.C. statistics (only with Q.C. smart card)
This Q.C. sub-menu allows the operator to print out all the stored cumulative data for all 3 levels on
the control blood. The information printed out will contain all the necessary statistical data for each
level. The levels can be selected individually or all at once.
It is highly recommended to print out all Quality Control Statistical data at the end of
each month for hard copy verification of control data.
Each File printout contains the following information: File Name (Blood Level), Lot # of the Control,
Expiration date of the control, Date and Time of the print data request, Date and Time of each control
run, Operator and Parameter results of each control run, the Reference Assay Mean, Upper, and Lower
limits, the Actual Mean results of the total control runs, the Standard Deviation value, and the Per-
cent Coefficient of Variation.
Printing Statistics:
Selecting Statistics
Main Menu
2- QC
4- Statistics. Press «Enter» key
Use the «Up» & «Down» arrow key to select one of the 3 levels to be printed out
Or select «ALL» to print out the 3 levels (a star * shows the current selection)
Press «Enter» key to print out the statistical data