Calibration Coefficient Limits - HORIBA ABX Micros 60 User Manual

Hematology analyzer
Table of Contents

Advertisement

PLT
MPV
RDW COEFF.
PDW COEFF.
PCT and PDW are not available in the United States. These parameters are strictly used
for research and investigational purposes only.
2.2.13.1. Print Coefficients
Printing calibration coefficients
Main menu
3- Calibration
2- Coefficients
2- Print Coeff.
The current calibration coefficients will automatically printout as indicated:
DATE : 01/20/2002
CURRENT
RDW COEFF
PDW COEFF
Fig.4-6: Calibration Coefficients printout

2.2.14. Calibration Coefficient limits

After any calibration has been performed on the
calibration coefficients are within their specified ranges as indicated:
Tab.4-2: Calibration Coefficient limits
Parameters
WBC
Coefficient limits
Minimum
0.89
Target
1.09
Maximum
1.29
If any of the calibration coefficients are out of their specified ranges after calibrating the
60, contact your local HORIBA ABX Technical Support Representative.
cros
COEFFICIENTS
TIME : 14:26
WBC
RBC
HGB
HCT
0.97
0.88
1.13
1.08
: 1.00
: 1.00
ABX Micros
RBC
HGB
HCT
PLT
0.73
0.83
0.87
0.99
0.89
1.11
1.08
1.20
1.05
1.39
1.29
1.41
ABX Micros
Quality
Assurance
PLT
MPV
0.95
0.92
60, Verify that all parameter
MPV
RDW
PDW
0.75
0.75
0.75
0.94
1.00
1.00
1.13
1.25
1.25
ABX Mi-
60
- User Manual - RAB042FEN - 19
Calibration