2.2.9. Calibration Failed

In order for
which contain 2 conditions.
1-
The Coefficient of Variations are out of thier specified limits as indicated in
of variation limits, page
2-
The difference between the «Target value» and the «Mean» for each parameter that failed cali-
bration, is greater than 20%.
When
ABX Micros
efficients are «Rejected», and the previous coefficients will remain unchanged in memory and the
screen will indicate: «CALIBRATION FAILED !»
When the Calibration STATUS indicates «FAILED» on one or more parameters, even
though stating «OK» on the other parameters, Calibration will not take place.
When the calibration fails, the operator may restart the calibration again or call your
local HORIBA ABX Technical Support representative for further instructions.
On the screen display: Rejected and saved coefficients can be displayed by using the «Up» and
«Down» arrow keys on the front panel. Press the «Esc» key to return to Main Menu.
«Failed» Calibration printout is as follows:
DATE : 01/10/2006
CALIBRATION FAILED
OPERATOR : ABC
LOT # : MCAL212
RUN
1 P
2
3
4
5
6
MEAN
TARGET
CV
% CHG
REJ. COEFF
CURRENT
STATUS
Fig.4-5: Failed Calibration printout
16 - RAB042FEN - User Manual -
60
ABX Micros
60
to «Fail» Calibration, the data must meet the statistical criteria
15".
60
«Fails calibration», the results will be printed out, the Calibration co-
CALIBRATION
TIME : 09:48
WBC
RBC
10.2
4.50
9.9
4.45
9.7
4.41
10.0
4.51
9.9
3.65
9.8
4.38
9.9
4.32
9.9
4.54
1.0
7.7
0.0
5.09
1.09
0.89
1.09
1.05
OK
FAILED
ABX Micros
HGB
HCT
PLT
14.0
38.0
242
14.0
37.4
246
14.0
37.2
237
14.2
37.9
251
14.1
31.8
254
14.1
36.8
248
14.1
36.5
247
13.5
37.2
260
0.7
6.3
2.3
5.74
1.92
5.26
1.11
1.08
1.20
1.07
1.02
1.26
OK
FAILED
OK
60
"Tab.4-1: Coefficient
MPV
7.6
7.9
7.7
7.7
7.5
7.5
7.7
7.7
1.8
0.0
0.94
0.94
OK