1.8. Extended Concentrated Cleaning

Concerns
◆RBC/PLT Chamber and Aperture cleaning
◆WBC/HGB Chamber and Aperture cleaning
Required Tools
5ml or 10ml syringe
Required Products
ABX MINOCLAIR 0.5L (P/n: 0401005), or bleach containing 12% active chlorine
Intervention Time
Approximately 30 minutes
Specific Kits or Consumable products
None
Procedure
1-
Run a backflush sequence
Running Backflush
Main menu
4- Service
1- Backflush
1-
Perform a Backflush cycle 2 more times.
2-
Drain the chambers
Draining chambers
Main menu
4- Service
2- Drain chambers
3-
Open the main door of the
4-
Locate the opening on the top of the chamber (shield or not)
5-
Fill both chambers with 5ml of the ABX MINOCLAIR.
6-
Let the solution set for 10 minutes in the chambers.
7-
After 10 minutes, run a Concentrated cleaning as described in
page
11".
8-
Once the cleaning cycle is complete, perform a Backflush cycle (see
quence, page
9-
Perform the Backflush (2) more times.
10-
Press the «Esc» key until you return to the Main Menu.
11-
Press the «Startup» key to run a startup. Verify that Background results are within their limits.
12-
Re-run Quality Control and verify that the results are within the Assay limits listed on the control
Assay sheet that comes with the controls.
13-
If the problem still persists, repeat steps
16 - RAB042FEN - User Manual -
60