Precision Claims - HORIBA ABX Micros 60 User Manual

Hematology analyzer
Table of Contents

Advertisement

Tab.1-6:
MINOTROL 16
Parameter
Control
M243 High
WBC
M243 Normal
M243 Low
M243 High
RBC
M243 Normal
M243 Low
M243 High
HGB
M243 Normal
M243 Low
M243 High
HCT
M243 Normal
M243 Low
M243 High
PLT
M243 Normal
M243 Low
Evaluation of Precision Performance of Clinical Chemistry Devices; Approved Guidelines, NCCLS doc-
ument EP-5 (ISBN 1-56238-145-8) 1999
*Source 510K submission # K030799

3.2. Precision Claims*

Tab.1-7:
Parameters
WBC
RBC
HGB
HCT
PLT
*Source 510K submission # K030799
Within run
CV% of run
CV%
means
1.19
1.01
1.39
1.11
3.06
1.68
0.9
0.62
1.02
0.88
1.46
1.07
1.72
1.28
0.85
1.17
1.49
3.5
1.06
1.62
1.51
1.38
1.6
1.6
2.89
1.37
9.17
7.23
5.14
6.87
%CV
<2.5%
<2.0%
<1.5%
<2.0%
<5.0%
ABX Micros
Specifications
Summary of Performance Data
CV% of Daily
Total impreci-
means
sion (CV%)
0.96
1.46
1.46
1.93
1.93
3.13
0.78
1.1
0.8
1.24
1.15
1.72
2.23
2.7
3.17
3.33
4.65
5.37
1.96
2.39
2
2.47
1.96
2.53
1.91
2.96
5.87
10.13
6.70
9.04
Nominal Values
3
10.0 x 10
/µL
6
5 x 10
/µL
15 g /dL
45 %
3
300 x 10
/µL
60
- User Manual - RAB042FEN - 9