1.2.4. Diluent & Waste connection
Connect the diluent container using the diluent straw and a 3 x 6 cristal tube (1 meter maximum)
on the diluent input located at the bottom of the instrument rear panel.
Connect the waste container using the cristal tube 3 x 6 on the waste output, and place the waste
container below the instrument level (under the bench).
Fig.6-7: Rear connections
Always follow the recommended procedures for waste disposal. Never connect the in-
strument wastes directly to the laboratory drain pipes.

1.2.5. Reagent priming

1.2.5.1. Bottles and containers set
Step to
"1.3.5. Prime Reagents, page 10"
1.2.5.2. Reagent pack
Step to
"1.3.5. Prime Reagents, page 10"
It is recommended not to remove the pack several times before the reagents are totally
used in order to avoid leak problems.
Maintenance
to run the priming of the replaced reagent.
to run the priming after the pack replacement.