2.2.7. Verify Calibration

1-
Once the calibration is complete and has Passed the Calibration criteria, Press the «Esc» key until
you have returned to the Main Menu.
2-
Run the remaining Calibrator material 3 times as a regular patient analysis. (Remember to gently
and thoroughly mix the material between each sample run.)
3-
When each cycle is complete, record the results. Calculate the Total and Mean values for each
parameter listed. Now compare the Mean value for each parameter to the Assay Mean values and
ranges listed for the
specific parameter Ranges on the Assay sheet. If all parameters are within their specified ranges,
Calibration is complete.
4-
Run Quality Control and verify that all 3 levels of control results are within their specified ranges.
Verify that all control parameter results are without Flags (H, L, *, $, and ! for HGB).