Troubleshooting System Operations - HORIBA ABX Micros 60 User Manual

Hematology analyzer
Table of Contents

Advertisement

2.4. Troubleshooting System Operations

Tab.6-5:
Problem Noted
No Power to the unit
No Display on LCD screen
Motor Errors on any motor
Maintenance
Possible Cause
Unit is powered Off and/or power
cord is disconnected from power
source
Possible blown power fuses
Possible disconnection of cable to
display
Possible incorrect adjustment on LCD
contrast
Possible Motor failure
Possible disconnected cable to motor
on main board
ABX Micros
& Troubleshooting
Troubleshooting
Corrective Action
Power On unit and/or plug in power
cord to power source
Call HORIBA ABX Technical support for
assistance
Call HORIBA ABX Technical support for
instructions on re-connection of cable
Press DEL key and Period ( . ) key si-
multaneously, re-adjust contrast
Check motor operation using
"1.3.7. Mechanical Checks, page 12"
Remove main cover of the unit, check
all connectors, call HORIBA ABX Tech-
nical support for assistance
60
- User Manual - RAB042FEN - 23