5. Labels

5.1. Main power label
In order to replace the 2x1A Fuses located under the power plug connection on the back of the an-
alyzer, carry out the following procedure:
- Do Not remove the instrument protection cover.
- Power OFF the analyzer.
- Disconnect the main power cable from the back power cord receptacle on the analyzer.
- Pull the little flap marked «250V fuse».
- Remove the fuses from their holding receptacle
- Check for the correct ohms on each fuse.
- Use only «Slow-blow» internal fuses.
- Use only fuses having the following characteristics:
for 100/120Vac supply:1A 250V SB
for 220/240Vac supply:1A 250V SB
5.2. Input/Output Label
Waste: Connect the waste output line to the (Waste position) fitting. Note the Waste label for Waste
output «Only»
Diluent: Connect the Diluent input line to the (Diluent position) fitting. Note the Diluent label for
Diluent input «Only»
Lyse: Connect the Clear tubing marked with a «MINILYSE» label to a straw and place it into the Lyse
reagent container.
Miniclean: Connect the Blue tubing marked with a «MINICLEAN» label to a straw and place it into
the Miniclean reagent container.
RS-232 output connection: Used only by HORIBA ABX qualified Engineers.
Printer connection: Do not connect any printer which has not been recommended by a HORIBA ABX
qualified Engineer.
5.3. «Biological risks» label
This label is located on the left side cover
Fig. 1: Biological risks
ABX Micros
60
- User Manual - RAB042FEN - 9