3. Summary of Performance Data

3.1. Precision (Reproducibility)*

The
ABX Micros
Date: 11-05-2002).
Three levels of ABX Minotrol 16 control material (Lot N°: M243) were run in duplicate twice daily
for 20 days. The results were used to quantify within run precision, SD of the run means, SD of the
daily means, and the Total Imprecision with the NCCLS EP-5 Guidelines.
.
Tab.1-5:
Parameter
WBC
RBC
HGB
HCT
PLT
8 - RAB042FEN - User Manual -
60
60
was initially calibrated with Minocal Calibrator (Lot N° MCAL 325 Expiry
MINOTROL 16
Within Run
Control
SD
0.23
M243 High
0.10
M243 Normal
0.06
M243 Low
0.05
M243 High
0.05
M243 Normal
0.04
M243 Low
0.31
M243 High
0.12
M243 Normal
0.09
M243 Low
0.52
M243 High
0.56
M243 Normal
0.26
M243 Low
13.71
M243 High
9.17
M243 Normal
5.83
M243 Low
ABX Micros
60
SD of Runs
SD of Daily
Means
Means
0.20
0.19
0.08
0.11
0.03
0.04
0.04
0.04
0.04
0.04
0.03
0.03
0.23
0.40
0.16
0.43
0.22
0.29
0.79
0.96
0.51
0.74
0.26
0.32
6.49
9.07
7.23
5.87
3.78
3.69
Total impreci-
sion (SD)
0.29
0.14
0.06
0.06
0.06
0.04
0.49
0.45
0.33
1.17
0.91
0.42
14.04
10.13
4.97