1.2.4. Running Control blood
3-
Verify that the lot number on the screen matches the lot # on the Commercial control blood.
4-
Once the Control blood has equilibrated to room temperature, Gently and thoroughly mix the lev-
el. Follow the product instructions that come with the control kit for proper mixing.
5-
Remove the cap and place the vial beneath the sample probe, raise the vial upwards so that the
sample probe enters into the blood.
6-
Press the «Start» key (or the Manual Sample Bar) on the front panel to start the cycle.