2.2.8. Calibration Passed

In order for
ABX Micros
which contain 2 conditions.
1-
The Coefficient of Variations must be within their limits as indicated
ation limits, page
15".
2-
The difference between the «Target value» and the «Mean» for each parameter calibrated, must
be less than 20%.
Tab.4-1: Coefficient of variation limits
Parameters
WBC
CV%
< 2.5
< 2.0
«Passed» calibration is as follows:
DATE : 01/10/2006
OPERATOR : ABC
LOT # : MCAL212
RUN
WBC
1 P
10.2
2
9.9
3
9.7
4
10.0
5
9.9
6
9.8
MEAN
9.9
TARGET
9.9
CV
1.0
% CHG
0.0
OLD CAL.
1.09
CURRENT
1.09
STATUS
OK
Fig.4-4: Passed Calibration printout
The «P» to the right of RUN # 1 indicates that the first calibration sample is not includ-
ed in the statistical calculations. This first sample is considered as a Calibrator material
prime.
The instrument wait for the operator to press any key to return to the MAIN MENU.
60
to «Pass» Calibration, the data must meet the statistical criteria
RBC
HGB
HCT
< 1.5
< 2.0
CALIBRATION
TIME : 09:48
RBC
HGB
4.50
14.0
4.45
14.0
4.41
14.0
4.51
14.2
4.43
14.1
4.38
14.1
4.44
14.1
4.54
13.5
1.0
0.7
2.25
5.74
0.89
1.11
0.91
1.07
OK
OK
ABX Micros
Quality
Assurance
"Tab.4-1: Coefficient of vari-
PLT
MPV
< 5.0
< 3.0
HCT
PLT
MPV
38.0
242
7.6
37.4
246
7.9
37.2
237
7.7
37.9
251
7.7
37.2
254
7.5
36.8
248
7.5
37.3
247
7.7
37.2
260
7.7
1.0
2.3
1.8
0.73
5.26
0.0
1.08
1.20
0.94
1.08
1.26
0.94
OK
OK
OK
60
- User Manual - RAB042FEN - 15
Calibration