Troubleshooting Specific Error Messages - HORIBA ABX Micros 60 User Manual

Hematology analyzer
Table of Contents

Advertisement

ABX Micros

2.5. Troubleshooting Specific ERROR Messages

Tab.6-6:
Error message
Startup Failed
Startup Failed. HGB reference failed.
Pressure syringe motor error, «Pack» units
Pressure syringe motor error, «separate reagent» units
Liquid syringe motor error
Carriage motor error
Needle motor error
Piercing motor error
24 - RAB042FEN - User Manual -
60
Corrective Action
Check and/or replace reagent(s)
Perform a concentrated cleaning
Cleaning, page
Remove and flush out sample probe Section 6,
1.7. Manual Probe Cleaning
startup
Check syringe block for leakage and call HORIBA ABX
Technical support
Possible electronic interferences. Call HORIBA ABX
Technical support for assistance
Check and/or replace reagent(s)
Perform a concentrated cleaning
Cleaning, page
Possible bad HGB blank reference voltage. Call HORIBA
ABX Technical support for assistance
For Pack units, remove Waste line from pack, push down
on and hold in pack waste valve, push on side of pack
to force air out of waste badder. Reconnect waste line.
Check for obstruction or restriction of Waste line
Waste container must be vented
Possible defective motor, Call HORIBA ABX Technical
support for assistance.
Check syringe block moving smoothly, call HORIBA ABX
Technical support
Possible disconnected cable from main board, remove
unit cover, check all connectors.
Possible defective motor, call HORIBA ABX Technical
support
Possible restriction of movement from the chambers to
home position, check for restriction
Possible defective carriage sensor. Proceed as described
in
"1.3.7.1. Check sensors test, page 12"
Possible defective motor, call HORIBA ABX Technical
support
Possible restriction of movement from the Down posi-
tion to Up home position, check for restriction
Possible defective needle sensor.Proceed as described
in
"1.3.7.1. Check sensors test, page 12"
Possible defective motor, call HORIBA ABX Technical
support
Possible Micro-switch failure behind Piercing mecha-
nism, check switch contact with applicator stick, call
HORIBA ABX Technical support for assistance
Restriction of movement to Up piercing position, check
for restriction
Possible defective piercing motor, call HORIBA ABX
Technical support
ABX Micros
60
"1.3.6. Concentrated
11", re-run startup
, page 6-14. Re-run
"1.3.6. Concentrated
11", re-run startup