Qc Results; Accepting Or Rejecting Qc Results - HORIBA ABX Micros 60 User Manual

Hematology analyzer
Table of Contents

Advertisement

1.2.5. QC results

When the control analysis cycle is complete, the results are displayed.
QC results review
Use the «Up» and «Down» arrow keys to scroll through the results.
The control results are printed out as indicated:
DATE : 01/20/2006
LOT # : M211
EXP DATE : 03/20/2006
OPERATOR : OP. 1
SEQ. # : 1
STARTUP PASSED
3
3
WBC :
2.3 10
/mm
6
3
RBC :
2.36 10
/mm
HGB :
5.6 L g/dl
HCT :
15.9 %
3
3
PLT :
70 10
/ mm
DIFF :
%LYM :
59.6 %
%MON :
14.2 %
%GRA :
26.2 %
WBC
50
100
250
300
450
Fig.4-1: QC results printout
Printed results may vary on the amount of data displayed on the printout. Limits, LMG's,
histograms, and parameters are all dependant upon the initial instrument setup. See
Section 5,
2.7. Print LMG Results

1.2.6. Accepting or Rejecting QC Results

The results from the control blood are compared to Assayed ranges stored on the Quality Control
smart card:
◆If any of the parameter results are «Out of Range», an «H»(High) or «L»(Low) will be shown
on the display and on the printout as well.
QC
LOW
Q.C. : 02
(1.7 - 2.5)
MCV :
(2.27 - 2.57)
MCH :
(5.7 - 6.7)
MCHC :
(14.2 - 18.2)
RDW :
(52 - 92)
MPV :
(52.7 - 66.7)
#LYM :
(7.4 - 19.4)
#MON :
(19.9 - 33.9)
#GRA :
RBC
30
100
230
, page 5-6S.
ABX Micros
Quality
Assurance
Quality Control
TIME : 09:25
3
67 µm
(63 - 71)
23.6 Pg
(23.6 - 27.6)
35.2 g/dl
(35.2 - 41.2)
12.7 %
(9.9 - 15.9)
3
(7.4 - 11.4)
9.4 µm
3
3
1.3 10
/ mm
(0.9 - 1.7)
3
3
(0.1 - 0.5)
0.3 10
/ mm
3
3
(0.2 - 1.0)
0.7 10
/ mm
PLT
5
10
20
30
60
- User Manual - RAB042FEN - 5