Dip Switch - EBARA EV-S20 Instruction Manual

Multi-stage dry vacuum pump ce / semi / nrtl model 200-220v(50/60hz)
Table of Contents

Advertisement

5.4 Dip Switch

Set the dip switches to select the operating modes as shown in Table 5.4 / 5.5 / 5.6.
No.
1
Data length
Cooling water &
2
N2 monitoring
3
Operation switched to
4
5
6
7
Dil N2 mode
8
BP start mode
No.
1
MP Speed control
2
BP Speed control
3
4
5
6
Remote Interface (IF)
7
8
LCD screen initialize
No.
1
Outputs the pump N2
2
3
4
5
6
7
8
DIP SW-A. No.1
PM10U
Table 5.4 Dip Switch-A Settings
Mode
Buzzer
According to signal
Remote
-------
-------
Standard mode
Table 5.5 Dip Switch-B Settings
Mode
-------
-------
-------
Exclusive special IF
-------
Carry out initialize
Table 5.6 Dip Switch-C Settings
Mode
warning.
-------
-------
-------
-------
-------
-------
-------
This switch allows you to select the data length out of 7 or 8
bits for the pump running status monitoring with the RS232C
communication port.
Off
7 bits
Constant
Only during operation
Not used
-------
-------
Automatic
Off
Not used
Not used
-------
-------
-------
-------
Off
Normal Open
-------
-------
-------
-------
-------
-------
-------
On
8 bits
Used
PUMP STOP
-------
-------
Dil N2-0 mode
Manual
On
Used
Used
-------
-------
-------
No use / standard IF
-------
Do not initialize
On
Normal Close
-------
-------
-------
-------
-------
-------
-------
EBARA
P.67
CORPORATION

Hide quick links:

Advertisement

Table of Contents
loading