(6)
Put the LCD controller on the upper part of the pump. Fix it with the tape.
(7)
Wrap the pump in a vinyl sheet.
(8)
Use the lifting brackets provided on the pump for slinging the pump to load
and unload. Fasten lifting brackets completely and push in until flush with
the seating surface. For sling, use a wire with a length so that the slinging
angle (that is, the angled subtended by the two wires) is within 60 degrees.
DANGER Do not enter the zone underneath the suspended pump.
WARNING For lifting the pump, use only qualified operator personnel.
(9)
When options such as an interface box are attached to the pump, be careful
to avoid damage due to contact by the wire rope.
(10) For transportation, secure the pump by lowering the adjustment feet. Place a
protective cloth around the pump to avoid shock and position protective
members between the outer cover and the wires in order to distribute the
load of the fastening wires.
PM10U
Be sure that the wire rope and crane used for lifting the pump are
in proper order and match the weight of the pump.
To prevent unequal weight distribution, suspend the pump by
ensuring that the slinging angle remains symmetrically centered.
Fig. 8.1 Slinging the Pump