Backing Up The Gauge Configuration - Emerson Rosemount 5900S Series Reference Manual

Radar level gauge
Hide thumbs Also See for Rosemount 5900S Series:
Table of Contents

Advertisement

Rosemount 5900S Series
6.2.2

Backing Up the Gauge Configuration

6-4
Input and holding registers of the Rosemount 5900S Radar Level Gauge can
be stored on disk. This can be useful for backup purposes and
troubleshooting. You can save a predefined set of holding registers to make a
backup copy of the current gauge configuration. The backup file can be used
to restore configuration of the 5900S at a later stage.
Single device
To save the current configuration to file for a single 5900S do the following:
1. Start the TankMaster WinSetup program.
2. In the WinSetup workspace window, click the right mouse button on the
5900S gauge icon.
3. Choose the Devices/Save Database to File option.
This option is also available from the Service/Devices menu.
4. Choose the Holding Registers and Predefined Registers options (the
User-Defined option should only be used for advanced service).
5. Click the Browse button, select a folder and type a name for the backup
file.
6. Click the Save button to start saving the database registers.
Reference Manual
00809-0100-5900, Rev CA
Section 6. Service and Troubleshooting
June 2014

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents