Device Status - Emerson Rosemount 5900S Series Reference Manual

Radar level gauge
Hide thumbs Also See for Rosemount 5900S Series:
Table of Contents

Advertisement

Reference Manual
00809-0100-5900, Rev CA
June 2014
6.3.1

Device Status

Table 6-2. Device status
messages
Running Boot Software
Device Warning
Device Error
BOOT Beta Version
APPL Beta Version
Level correction error
Invalid Measurement
Write Protected
Default Database
Simulation Active
RM Reprogramming In Progress
Section 6. Service and Troubleshooting
Device Status messages that may appear on the display of the Rosemount
2410 Tank Hub, or in the Rosemount TankMaster program, are shown in
Table 6-2. Device Status can be found in Input register 4000. See "Viewing
Input and Holding Registers" on page 6-2 for more information on how to view
Input registers.
Message
• The application software could not
be started.
• The application SW is not loaded in
the flash memory
• The previous upload of the flash
SW failed
A device warning is active.
A device error is active.
Beta version of boot program used
Beta version of application program
used
The LPG module is enabled but the
module is either incorrectly
configured, or there is no sensor input
data for pressure or temperature.
The level gauge indicates that
measurement is invalid. This may be
caused by an actual measurement
problem or some other error
indication.
The configuration registers are write
protected.
All configuration registers are set to
default values.
The 5900S is in simulation mode
New software is downloaded to the
5900S Radar Level Gauge
Description
Reprogram the gauge with new software.
Contact Emerson Process
Management/Rosemount TankGauging
service department.
See "Warning Messages" on page 6-20 for
details.
See "Error Messages" on page 6-21 for
details.
Ensure that approved software is used
Ensure that approved software is used
See Input register 4702
LPGIregArea-LPG_Corr_Error for more
information.
Check Error Messages, Warning Messages
and Measurement Status for details.
Do one of the following:
1. Use the Lock/Unlock function to turn off the
software write protection.
2. Change position of the Write Protection
switch to OFF.
Ensure that device calibration is valid.
Reset 5900S simulation mode.
Verify 5900S operation when reprogramming
is finished
Rosemount 5900S Series
Action
6-19

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents