Emerson Rosemount 5900S Series Reference Manual page 137

Radar level gauge
Hide thumbs Also See for Rosemount 5900S Series:
Table of Contents

Advertisement

Reference Manual
00809-0100-5900, Rev CA
June 2014
Section 4. Configuration
MAINT_ALARMS
A maintenance alarm indicates that the device or some part of the device
needs maintenance soon. If the condition is ignored, the device will eventually
fail. There are five parameters associated with MAINT_ALARMS, they are
described below.
MAINT_ENABLE
The MAINT_ENABLE parameter contains a list of conditions indicating
that the device or some part of the device needs maintenance soon. Below
is a list of the conditions in order of priority from 1 to 2, where 2 is the
highest priority. This priority is not the same as the MAINT_PRI parameter
described below. It is hard coded within the device and is not user
configurable.
Below is a list of the conditions
1. Device Major Information
2. Device Warning
MAINT_MASK
The MAINT_MASK parameter will mask any of the failed conditions listed
in MAINT_ENABLED. A bit on means that the condition is masked out
from alarming and will not be reported.
MAINT_PRI
MAINT_PRI designates the alarming priority of the MAINT_ALM, see
"Alarm Priority" on page 4-46. The default is 0 and the recommended
values are 3 to 7.
MAINT_ACTIVE
The MAINT_ACTIVE parameter displays which of the alarms are active.
MAINT_ALM
An alarm indicating that the device needs maintenance soon. If the
condition is ignored, the device will eventually fail.
(1) Note that maintenance alarms are not enabled by default.
Rosemount 5900S Series
(1)
:
4-53

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents