ADVISE_ALARMS
An advisory alarm indicates informative conditions that do not have a direct
impact on the primary functions of the device. There are five parameters
associated with ADVISE_ALARMS, they are described below.
ADVISE_ENABLE
The ADVISE_ENABLE parameter contains a list of informative conditions
that do not have a direct impact on the primary functions of the device.
Below is a list of the conditions in order of priority from 1 to 2, where 2 is
the highest priority. This priority is not the same as the ADVISE_PRI
parameter described below. It is hard coded within the device and is not
user configurable.
Below is a list of the advisories
1. Device Minor Information
2. PlantWeb Alerts Simulation Active
ADVISE_MASK
The ADVISE_MASK parameter will mask any of the failed conditions listed
in ADVISE_ENABLE. A bit on means the condition is masked out from
alarming and will not be reported.
ADVISE_PRI
ADVISE_PRI designates the alarming priority of the ADVISE_ALM, see
"Alarm Priority" on page 4-46. The default is 0 and the recommended
values are 1 or 2.
ADVISE_ACTIVE
The ADVISE_ACTIVE parameter displays which of the advisories is
active.
ADVISE_ALM
An alarm that indicates advisory alarms. These conditions do not have a
direct impact on the process or device integrity.
Recommended Actions for PlantWeb Alerts
The RECOMMENDED_ACTION parameter displays a text string that will give
a recommended course of action to take based on which type and which
specific event of the PlantWeb alerts is active (See Table 6-10 on page 6-29).
(1) Note that advisory alarms are not enabled by default.