Rosemount 5900S Series
6.3.2
Warning
Messages
Table 6-3. Warning messages
RAM warning
FPROM warning
HREG warning
Other memory warning
MWM warning
RM warning
Other hardware warning
Configuration warning
SW warning
6-20
Table 6-3 shows a list of Warning messages that may appear on the integral
display of the Rosemount 2410 Tank Hub and in the Rosemount TankMaster
program. You also have the option to view Input Register 1004 for an
overview of active device warnings. Warnings are less serious than errors.
For each warning message that may appear, detailed information can be
found in Input registers 6100 to 6130 as shown in Table 6-3.
Message
Input register no. 6100.
Bit 0: DSP Stack
Bit 1: DSP RAM low
Input register no. 6102.
Input register no. 6104.
Bit 0: DSP Factory holding registers
Input register no. 6106.
Input register no. 6108.
Bit 1: Version mismatch between PM
and RM
Input register no. 6110
Bit 1: SW config
Bit 5: FPROM Checksum
Bit 6: FPROM Version
Bit 9: HREG Checksum
Bit 10: HREG Limit
Bit 11: HREG Write
Bit 12: HREG Read
Bit 13: HREG Version
Bit 14: MWM Invalid Id
Bit 30: SW Serious Warning
Input register no. 6122.
Input register no. 6128.
Bit 0: Super Test Active
Bit 1: ATP Table Invalid
Bit 2: Special Correction Table Invalid
Bit 3: Near Zone Correction Table
Invalid
Bit 4: Config Model Code invalid
Bit 5: Config LPG Pins Visible
Bit 6: Config LPG Error
Bit 7: Simulation Mode Used
Bit 8: Default Sweep Mode Used
Bit 9: Test Sweep used
Bit 10: ACT Table Invalid
Bit 11: UCT Table Invalid
Bit 12: Simple Simulation Mode
Warning
Bit 13: Ramp Simulation Mode
Warning
Bit 14: TSM Filter Too Narrow
Bit 15: MMS Offset Update disabled
Input register no. 6130.
Bit 8: DSP Undefined software
warning
Description
Contact Emerson Process
Management/Rosemount TankGauging
service department.
Load default database and restart the 5900S.
Contact Emerson Process
Management/Rosemount TankGauging
service department if the problem persists.
Contact Emerson Process
Management/Rosemount TankGauging
service department
• Load the default database and restart the
• Configure the level gauge or load a backup
• Contact Rosemount TankGauging service
Contact Emerson Process
Management/Rosemount TankGauging
service department
Section 6. Service and Troubleshooting
Reference Manual
00809-0100-5900, Rev CA
Action
level gauge, see "Loading the Default
Database" on page 6-14.
configuration file (see "To Recover a
Backup Configuration Database" on
page 6-6).
department if the problem persists.