Procedure
Follow these instructions when you make hand dip measurements:
•
hand dip until three consecutive readings within 1 mm are obtained
•
correct the tape according to the calibration record
•
note the hand dipped ullage and the gauge level reading
simultaneously
Do not calibrate when
•
the tank is being emptied or filled
•
agitators are running
•
when there are windy conditions
•
when there is foam on the product surface