Rosemount 5900S Series
4-62
20. Select check boxes that correspond to the conditions in the tank. Use as
few options as possible. It is recommended that no more than two
options are used simultaneously.
21. Choose Product Dielectric Range from the drop-down list. Use the
Unknown option if the correct value range is unknown or if the contents
of the tank is changing on a regular basis.
F
Fieldbus parameters:
OUNDATION
TRANSDUCER 1100>PRODUCT_DC
TRANSDUCER 1100>TANK_ENVIRONMENT
22. Click the Finish button.
Reference Manual
00809-0100-5900, Rev CA
June 2014
Section 4. Configuration