Repairing the Flow Cells

When required
Tools required
Parts required
Preparations
The shown cell parts will differ depending upon the flow cell type. For detailed parts
NOTE
schematics, refer to above mentioned pages.
1 - Cell Screw
2 - Conical Springs
3 - Ring #1 PEEK
4 - Gasket #1
5 - Window Quartz
6 - Gasket #2
7 - Cell cover assembly
8 - Ring #2 PEEK
1100 Series Variable Wavelength Detector Reference Manual 91
Repairing the Variable Wavelength Detector
If the flow cell needs repair due to leaks or
contaminations.
Wrench 1/4 inch for capillary connections
Hexagonal wrench 4 mm
Tooth picks
See
"Standard Flow Cell (G1314-60086)"
See
"Standard Flow Cell (G1314-60080)"
See
"Micro Flow Cell"
on page 146.
See
"Semimicro Flow Cell"
See
"High Pressure Flow Cell"
Turn off the flow.
Remove the front cover.
Remove the flow cell, see
on page 142.
on page 144
on page 148.
on page 150.
"Exchanging a Flow Cell"
4
on page 91.