Exchanging Internal Parts - Agilent Technologies 1100 Series Reference Manual

Variable wavelength detector
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

4
Repairing the Variable Wavelength Detector
1 Remove the front cover to have access to the leak handling system.
2 Pull the leak funnel out of the leak funnel holder.
3 Pull the leak funnel with the tubing out of its location.
4 Replace the leak funnel and/or the tubing.
5 Insert the leak funnel with the tubing in its position.
6 Insert the leak funnel into the leak funnel holder.
7 Replace the front cover.
Figure 22
Replacing Waste Handling System Parts

Exchanging Internal Parts

The following procedures require opening the main cover of the detector.
Always
WARNING
ensure the detector is disconnected from the line power when the main cover is
removed. The security lever at the power input socket prevents the detector
cover from being taken off when line power is still connected.
To disconnect the detector from line, unplug the power cord. The power supply
still
98 1100 Series Variable Wavelength Detector Reference Manual