Setting the 8-bit Configuration Switch

The 8-bit configuration switch is located next to the GPIB connector. Switch settings
provide configuration parameters for GPIB address, serial communication protocol and
instrument specific initialization procedures.
Factory setting is
shown for the
variable wavelength
detector
Figure 52
Table 52
Switches 1 and 2 define which set of parameters (for example, for GPIB, RS-232C, and
so on) will be changed. Once the change has been completed, the instrument must be
powered up again in order to store the values in the non-volatile memory.
In the non-volatile memory, the parameters are kept, regardless of whether you turn the
instrument off and on again. They will be kept until the same set of parameters is
changed and the power is reset. All other previously stored configuration settings will
still remain in the non-volatile memory.
200 1100 Series Variable Wavelength Detector Reference Manual
8- bit Configuration Switch
8- bit Configuration Switch
Mode Select
1
GPIB
0
RS-232C
0
Reserved
1
TEST/BOOT
1
2
3
4
0
GPIB address
1
Baud rat e
0
Reserve d
1
RSVD
SYS
5
6
7
Data
Parity
bits
RSVD
RSVD
8
FC