xxxxis the product number, e.g. 1314 for the G1314A VWD), and vvvis
the revision number, for example 108 is revision 1.08 For instructions
refer to your user interface.
Update of main system can be done in the resident system only.
Update of the resident system can be done in the main system only.
NOTE
Figure 48

Raw Data Conversion to Absorbance

The raw data flow (from the photovariable wavelength) and conversion to absorbance
spectra with variance for each data point is a multiple step process. This process is
outlined in
1100 Series Variable Wavelength Detector Reference Manual 191
Resident System
Firmware Update Mechanism
Figure
49.
Introduction to the Variable Wavelength Dector
Main FW update
Resident FW update
Main System
6