4
Repairing the Variable Wavelength Detector
11 Reinstall the top foam section, top cover and front cover, see
12 Replace detector into the stack and reconnect the cables.
If a new VWM board was installed, update the serial number information of the
NOTE
detector in the user interface, see procedure below.
13 Check the firmware revision of the module. If the firmware revision is older than the
104 1100 Series Variable Wavelength Detector Reference Manual
and the Top Cover"
on page 131.
current firmware revision of the module, update the firmware using the standard
firmware update procedure, see
"Replacing the Detector's Firmware"
"Installing the Foam
on page 135.