Troubleshooting and Test Functions
• Defective power supply.
Suggested Actions
✔
✔
✔
✔
✔
Calibration Failed
The intensity maximum was not found during wavelength calibration.
Calibration 0 Failed: Zero-order calibration failed.
Calibration 1 Failed: 656 nm calibration failed.
During zero-order and 656 nm calibration, the detector searches for the intensity
maximum. If the maximum is not detected within the scan range, the error message is
generated.
Probable Causes
• Lamp is off.
• Incorrect flow cell installation.
• Flow cell contamination or air bubbles.
• Intensity too low.
• Current step value too far from maximum.
• Misalignment of the grating assembly.
• Defective grating assembly.
• Defective VWM board.
52 1100 Series Variable Wavelength Detector Reference Manual
Ensure the lamp is connected.
Replace the top foam, and turn on the lamp.
Exchange the lamp.
Exchange the VWM board.
Exchange the power supply.