J502 - Deuterium lamp
S1 - Configuration switch
B - Battery
Figure 24
Location of Connectors on VWM Board
7 Remove the processor board. Place the board on the ESD kit.
If you also have to replace other assemblies, continue with that work first.
NOTE
8 On the new board check the switch setting of address switch S1, see
An incorrect switch setting (for example, TEST/BOOT) may cause the module to turn in
a
NOTE
basic mode (yellow or red flashing status light). In such a case turn off the module, re-
set the address switches, and turn on the module again.
9 Install the new processor board and reconnect the connectors. Make sure that the
10 Refit the screws at the REMOTE and GPIB connectors and the nut of the analog
1100 Series Variable Wavelength Detector Reference Manual 103
J400
J401
Configuration Switch"
board is fitted correctly in the board recesses in the holes at the rear panel.
connector.
Repairing the Variable Wavelength Detector
on page 208.