Figure 11
Intensity Test
Intensity Test Evaluation (Agilent ChemStation only)
The Agilent ChemStation evaluates three values automatically and displays the limits
for each value, the average, the minimum and the maximum of all data points and
passed or failed for each value.
Table 5
Range
High intensity
Average intensity
Lowest intensity
Test Failed
Probable Causes
• Absorbing solvent in flow cell.
1100 Series Variable Wavelength Detector Reference Manual 65
Intensity Limits
Limit (counts)
> 10000
> 5000
> 200
Troubleshooting and Test Functions
3