Agilent Technologies 1100 Series Reference Manual page 219

Variable wavelength detector
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

Info Log
Use the m-key to receive a pop-up menu, Select Info Log. A list of the last events are
listed. For troubleshooting reasons they can be printed or saved to a file on the PCMCIA
card (using the m- key for the context sensitive menu ).
EMF (Early Maintenance Feedback)
Use the F1 key (EMF) to set EMF parameters. Choose menu item 1 (Setup limits) to
select lamp-on level for the UV lamp at which you want to receive a warning.
If a set limit has been exceeded, a message box will pop up. This will not stop a sequence
or run (information only to plan maintenance activities). If you press Reset, the limits
will be removed. Ignore will continue to keep the EMF flag set.
1100 Series Variable Wavelength Detector Reference Manual 217
Control Module Screens for the Agilent 1100 Variable Wavelength Detector
Use the F2 key (System Log) or F3 key (Error Log) to look for errors.