Flow Connections Of The Column Compartment - Agilent Technologies 1290 User Manual

Infinity thermostatted column compartment
Hide thumbs Also See for 1290:
Table of Contents

Advertisement

3
Installing the Module

Flow Connections of the Column Compartment

Flow Connections of the Column Compartment

Flow Connections Using Low Dispersion Heat Exchangers
Parts required
#
1
1
1
Preparations
• Install the column compartment
• Install the additional heat exchangers
Toxic, flammable and hazardous solvents, samples and reagents
WA R N I N G
The handling of solvents, samples and reagents can hold health and safety risks.
1
Press release buttons and remove front cover to gain
access to heater area.
60
Description
Other modules
Parts from Accessory Kit
Two wrenches 1/4 – 5/16 inch for capillary connections
When working with these substances observe appropriate safety procedures (for
example by wearing goggles, safety gloves and protective clothing) as described in
the material handling and safety data sheet supplied by the vendor, and follow good
laboratory practice.
The volume of substances should be reduced to the minimum required for the
analysis.
Do not operate the instrument in an explosive atmosphere.
2
Place the column on the left heat exchanger assembly
and connect the capillaries to the column.
1290 Infinity TCC User Manual