Removing the Optical Unit

When required
Tools required Screwdriver POZI 1 PT3
Parts required Optical unit G1314-69060 (exchange part)
Preparation for this procedure are:
Turn off the detector.
Disconnect the power cable.
Disconnect capillaries.
Remove the detector from stack and place it on the
working bench.
Remove the front cover, top cover and top foam
section, see
""Removing the Top Cover and Top
Foams"
on page 103.
2 Remove the lamp cover and the plate above the flow
1100 Series Variable Wavelength Detector Reference Manual 115
Repairing the Variable Wavelength Detector
When other assemblies have to be removed or when defective
3 Disconnect the deuterium lamp connector and all
cell.
cables coming from the optical unit to the main
board.