Repairing the Variable Wavelength Detector
Replace the detector into the stack and reconnect the cables and capillaries.
Turn on the detector.

Replacing the Interface Board

When
required
Part
required
Tools
required
1
128 1100 Series Variable Wavelength Detector Reference Manual
For all repairs inside the detector or for installation of the
board
Interface board (BCD) G1351-68701 with external contacts
and BCD outputs Interface board (LAN), see
Interface Boards"
None
Install the ESD strap.
Move the power lock
across the power
inlet.
Place the top part on the bench and insert
"Optional
on page 200.
the left and
2
If required, unscrew
and remove the
interface board.
Place the board on
the ESD kit.
2